Abstract

An evaluation of the accuracy in the application of the k0-INAA method using Zn foil as comparator is presented. A good agreement was found in the precision within analysts and between them, as well as in the assessment of trueness for most elements. The determination of important experimental parameters like gamma peak counting efficiency, γ-γ true coincidence, comparator preparation and quality assurance/quality control is also described and discussed. © 2010 Elsevier B.V. All rights reserved.
Original languageAmerican English
Pages419-424
Number of pages6
DOIs
StatePublished - 11 Oct 2010
Externally publishedYes
EventNuclear Instruments and Methods in Physics Research, Section A: Accelerators, Spectrometers, Detectors and Associated Equipment -
Duration: 11 Oct 2010 → …

Conference

ConferenceNuclear Instruments and Methods in Physics Research, Section A: Accelerators, Spectrometers, Detectors and Associated Equipment
Period11/10/10 → …

Fingerprint Dive into the research topics of 'K<inf>0</inf>-INAA method accuracy using Zn as comparator'. Together they form a unique fingerprint.

  • Cite this

    Bedregal, P., Mendoza, P., Ubillús, M., & Montoya, E. (2010). K<inf>0</inf>-INAA method accuracy using Zn as comparator. 419-424. Paper presented at Nuclear Instruments and Methods in Physics Research, Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, . https://doi.org/10.1016/j.nima.2010.01.043