TY - JOUR
T1 - K0-INAA method accuracy using Zn as comparator
AU - Bedregal, P.
AU - Mendoza, P.
AU - Ubillús, M.
AU - Montoya, E.
PY - 2010/10/11
Y1 - 2010/10/11
N2 - An evaluation of the accuracy in the application of the k0-INAA method using Zn foil as comparator is presented. A good agreement was found in the precision within analysts and between them, as well as in the assessment of trueness for most elements. The determination of important experimental parameters like gamma peak counting efficiency, γ-γ true coincidence, comparator preparation and quality assurance/quality control is also described and discussed.
AB - An evaluation of the accuracy in the application of the k0-INAA method using Zn foil as comparator is presented. A good agreement was found in the precision within analysts and between them, as well as in the assessment of trueness for most elements. The determination of important experimental parameters like gamma peak counting efficiency, γ-γ true coincidence, comparator preparation and quality assurance/quality control is also described and discussed.
KW - Gamma counting efficiency
KW - K-INAA
KW - Precision assessment
KW - Trueness assessment
KW - Zn comparator
UR - http://www.scopus.com/inward/record.url?scp=77956336947&partnerID=8YFLogxK
U2 - 10.1016/j.nima.2010.01.043
DO - 10.1016/j.nima.2010.01.043
M3 - Artículo
AN - SCOPUS:77956336947
SN - 0168-9002
VL - 622
SP - 419
EP - 424
JO - Nuclear Instruments and Methods in Physics Research, Section A: Accelerators, Spectrometers, Detectors and Associated Equipment
JF - Nuclear Instruments and Methods in Physics Research, Section A: Accelerators, Spectrometers, Detectors and Associated Equipment
IS - 2
ER -