TY - JOUR
T1 - Validation of an HPLC Methodology for the Identification and Quantification of Biogenic Amines in Chicken Meat
AU - Lázaro, César A.
AU - Conte-Júnior, Carlos A.
AU - Cunha, Fernanda L.
AU - Mársico, Eliane T.
AU - Mano, Sérgio B.
AU - Franco, Robson M.
N1 - Funding Information:
Acknowledgments César Lázaro acknowledges the Coordenação de Aperfeiçoamento de Pessoal de Nível Superior for the scholarship Programa Estudante Convenio Pos-Graduação.
PY - 2013/8
Y1 - 2013/8
N2 - This study validated a high performance liquid chromatography (HPLC) method to determine biogenic amines in chicken meat. For the identification of biogenic amines, an isocratic elution system coupled with a UV detector (254 nm) was used after a perchloric acid (5 %) extraction and benzoyl chloride derivatization of the samples. The standards of tyramine, putrescine, cadaverine, spermidine, and spermine were used for the following validation parameters: selectivity, linearity, accuracy, recovery, limit of detection and quantification, and robustness. Finally, chicken meat commercialized in two types of packaging was evaluated. The results showed an excellent selectivity and separation of all amines, r 2 > 0.99, relative standard deviation <5 %, recovery between 64 and 112 %, limits of detection and quantification between 0.03-1.25 and 0.15-5.00 μg L-1, respectively, and appropriate robustness for the proposed methodology. Moreover, both chicken meat commercial packages had similar values for all amines; only tyramine was significantly different (P ≤ 0.05). The proposed method was suitable to detection and quantification of simultaneous five biogenic amines in chicken meat.
AB - This study validated a high performance liquid chromatography (HPLC) method to determine biogenic amines in chicken meat. For the identification of biogenic amines, an isocratic elution system coupled with a UV detector (254 nm) was used after a perchloric acid (5 %) extraction and benzoyl chloride derivatization of the samples. The standards of tyramine, putrescine, cadaverine, spermidine, and spermine were used for the following validation parameters: selectivity, linearity, accuracy, recovery, limit of detection and quantification, and robustness. Finally, chicken meat commercialized in two types of packaging was evaluated. The results showed an excellent selectivity and separation of all amines, r 2 > 0.99, relative standard deviation <5 %, recovery between 64 and 112 %, limits of detection and quantification between 0.03-1.25 and 0.15-5.00 μg L-1, respectively, and appropriate robustness for the proposed methodology. Moreover, both chicken meat commercial packages had similar values for all amines; only tyramine was significantly different (P ≤ 0.05). The proposed method was suitable to detection and quantification of simultaneous five biogenic amines in chicken meat.
KW - Biogenic amines
KW - Chicken
KW - HPLC
KW - Meat
KW - Validation
UR - http://www.scopus.com/inward/record.url?scp=84879799616&partnerID=8YFLogxK
U2 - 10.1007/s12161-013-9565-0
DO - 10.1007/s12161-013-9565-0
M3 - Artículo
AN - SCOPUS:84879799616
SN - 1936-9751
VL - 6
SP - 1024
EP - 1032
JO - Food Analytical Methods
JF - Food Analytical Methods
IS - 4
ER -